Seel T, Schneider S, Affeld K, Schauer T (2015)
Publication Type: Journal article
Publication year: 2015
Book Volume: 63
Pages Range: 5-13
Journal Issue: 1
For a novel blood pressure measurement system, we design a cascaded controller that consists of two classic feedback loops and an iterative learning algorithm.We show that the bandwidth of the feedback loops is limited by the measurement noise of the ultrasound sensor and that this restriction can be circumvented by exploiting the repetitive nature of the disturbance that is induced by the oscillating arterial pressure.
APA:
Seel, T., Schneider, S., Affeld, K., & Schauer, T. (2015). Entwurf einer lernenden Kaskadenregelung für ein nichtinvasives kontinuierliches Blutdruckmesssystem. At-Automatisierungstechnik, 63(1), 5-13. https://doi.org/10.1515/auto-2014-1130
MLA:
Seel, Thomas, et al. "Entwurf einer lernenden Kaskadenregelung für ein nichtinvasives kontinuierliches Blutdruckmesssystem." At-Automatisierungstechnik 63.1 (2015): 5-13.
BibTeX: Download