Rothammer B, Neusser K, Bartz M, Wartzack S, Kistler M (2024)
Publication Language: English
Publication Type: Conference contribution, Conference Contribution
Publication year: 2024
Experimental in vitro testing of the implants was performed in a knee simulator according to ISO 14243-1. The wear mass of the tibial inlay was determined gravimetrically with an analytical balance. By coating both articulation partners, a reduction in the wear mass of ~57% compared to CoCr/CPE and ~62% compared to Ti64/CPE could be achieved.
APA:
Rothammer, B., Neusser, K., Bartz, M., Wartzack, S., & Kistler, M. (2024). A knee simulator pilot study: Performance of amorphous carbon coated articulating surfaces for total knee replacement. In Gesellschaft für Tribologie e.V. (Eds.), Proceedings of the 65. Tribologie-Fachtagung. Reibung, Schmierung und Verschleiß. Forschung und praktische Anwendungen.. Göttingen, DE.
MLA:
Rothammer, Benedict, et al. "A knee simulator pilot study: Performance of amorphous carbon coated articulating surfaces for total knee replacement." Proceedings of the 65. Tribologie-Fachtagung. Reibung, Schmierung und Verschleiß. Forschung und praktische Anwendungen., Göttingen Ed. Gesellschaft für Tribologie e.V., 2024.
BibTeX: Download