Nguyen HGT, Espinal L, Van Zee RD, Thommes M, Toman B, Hudson MSL, Mangano E, Brandani S, Broom DP, Benham MJ, Cychosz K, Bertier PG, Yang F, Krooss BM, Siegelman RL, Hakuman M, Nakai K, Ebner AD, Erden L, Ritter JA, Moran A, Talu O, Huang Y, Walton KS, Billemont P, De Weireld G (2018)
Publication Type: Journal article
Publication year: 2018
Book Volume: 24
Pages Range: 531-539
Journal Issue: 6
DOI: 10.1007/s10450-018-9958-x
This paper reports the results of an international interlaboratory study led by the National Institute of Standards and Technology (NIST) on the measurement of high-pressure surface excess carbon dioxide adsorption isotherms on NIST Reference Material RM 8852 (ammonium ZSM-5 zeolite), at 293.15 K (20 °C) from 1 kPa up to 4.5 MPa. Eleven laboratories participated in this exercise and, for the first time, high-pressure adsorption reference data are reported using a reference material. An empirical reference equation nex=d(1+exp[(-ln(P)+a)/b])c, [n
APA:
Nguyen, H.G.T., Espinal, L., Van Zee, R.D., Thommes, M., Toman, B., Hudson, M.S.L.,... De Weireld, G. (2018). A reference high-pressure CO2 adsorption isotherm for ammonium ZSM-5 zeolite: results of an interlaboratory study. Adsorption-Journal of the International Adsorption Society, 24(6), 531-539. https://dx.doi.org/10.1007/s10450-018-9958-x
MLA:
Nguyen, H. G. T., et al. "A reference high-pressure CO2 adsorption isotherm for ammonium ZSM-5 zeolite: results of an interlaboratory study." Adsorption-Journal of the International Adsorption Society 24.6 (2018): 531-539.
BibTeX: Download