Acceptance and reverification testing for industrial computed tomography - a simulative study on geometrical misalignments

Wohlgemuth F, Hausotte T, Schmidt I, Kimmig W, Imkamp KD (2022)


Publication Type: Journal article

Publication year: 2022

Journal

Book Volume: 11

Pages Range: 171-186

Journal Issue: 1

DOI: 10.5194/jsss-11-171-2022

Abstract

Acceptance and reverification testing for industrial X-ray computed tomography (CT) is described in different standards (E DIN EN ISO 10360-11:2021-)4, 2021; VDI/VDE 2630 Blatt 1.3, 2011; ASME B89.4.232020, 2020). The characterisation and testing of CT system performance are often achieved with test artefacts containing spheres. This simulative study characterises the influence of different geometrical error sources - or geometrical misalignments - on these sphere measurements. The two measurands on which this study focuses are the sphere centre-to-centre distances and the sphere probing form errors.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Wohlgemuth, F., Hausotte, T., Schmidt, I., Kimmig, W., & Imkamp, K.D. (2022). Acceptance and reverification testing for industrial computed tomography - a simulative study on geometrical misalignments. Journal of Sensors and Sensor Systems, 11(1), 171-186. https://dx.doi.org/10.5194/jsss-11-171-2022

MLA:

Wohlgemuth, Florian, et al. "Acceptance and reverification testing for industrial computed tomography - a simulative study on geometrical misalignments." Journal of Sensors and Sensor Systems 11.1 (2022): 171-186.

BibTeX: Download