Abstract

The determination of measurement uncertainties in dimensional X-ray computed tomography measurement technology is a complex task for which many different influences have to be taken into account. This work examines the influence of the scan mode in combination with the number of projections and an eccentric clamping on the quality of the measurement. It shows that the workpiece positioning has a considerable influence on the result in the case of continuous scan mode. In addition, an applicationdependent potential for reducing the lead times is described.