Kirchner J, Roth N, Meyer A, Fischer G (2016)
Publication Type: Conference contribution
Publication year: 2016
Conference Proceedings Title: Proc. IEEE Sensors Conf.
DOI: 10.1109/ICSENS.2016.7808736
The prototype of a capacitive ECG measurement device with purely passive signal transmission between electrodes and differential amplifier is presented. This design is intended to improve transmission symmetry and thus common-mode rejection. The feasibility of the approach is demonstrated by measurements with different materials between body surface and electrodes. The ECG module shall be incorporated in a mobile sensor platform in the form of a plaster for the monitoring of chronic diseases.
APA:
Kirchner, J., Roth, N., Meyer, A., & Fischer, G. (2016). ECG Measurement by Use of Passive Capacitively Coupled Electrodes. In Proc. IEEE Sensors Conf.. Orlando, FL, US.
MLA:
Kirchner, Jens, et al. "ECG Measurement by Use of Passive Capacitively Coupled Electrodes." Proceedings of the IEEE Sensors, Orlando, FL 2016.
BibTeX: Download