Cardiac T2 mapping: Robustness and homogeneity of standardized in-line analysis

Wiesmüller M, Wüst W, Heiß R, Treutlein C, Uder M, May M (2020)


Publication Type: Journal article

Publication year: 2020

Journal

Book Volume: 22

Article Number: 39

Journal Issue: 1

DOI: 10.1186/s12968-020-00619-x

Abstract

Background and purpose: Interpretation of T2 values remains difficult due to limited comparability across hardware and software systems and the lack of validated measurement recommendations for the number and orientation of mandatory slices. Our aims were to provide a standardized comparison of intra- and inter-individual T2 values in the short and long axes and to investigate inter-scanner reproducibility. Method and materials: Ninety cardiovascular magnetic resonance (CMR) studies in 30 healthy subjects were performed with three identical 1.5 T CMR scanners (same hardware and software) using a balanced steady-state free precession (bSSFP) gradient echo sequence in three short axis (SAx) and three long axis (LAx) views. A commercially available T2 mapping software package of the latest generation with automatic in-line motion correction was used for acquisition. Regions of interest were manually drawn in each of the 16 myocardial segments according to the American Heart Association (AHA) model in three SAx and three LAx acquisitions. Analysis of inter-scanner, inter-segmental, intra-segmental, inter-regional and inter-level differences was performed. Results: Inter-scanner reproducibility was high and the mean myocardial T2 value for all evaluated segments was 45.7 ± 3.4 ms. Significant inter-segmental variations of mean T2 values were found. Mean intra-segmental T2 values were comparable between LAx and SAx acquisitions in 72%. Significantly higher T2 values were found in apical segments and a significant disparity among different regions was found for SAx and LAx orientations. Conclusion: Standardized cardiac T2 mapping is highly reproducible on identical CMR systems. T2 values vary significantly between single heart segments, regions, levels, and axes in young, healthy subjects.

Authors with CRIS profile

How to cite

APA:

Wiesmüller, M., Wüst, W., Heiß, R., Treutlein, C., Uder, M., & May, M. (2020). Cardiac T2 mapping: Robustness and homogeneity of standardized in-line analysis. Journal of Cardiovascular Magnetic Resonance, 22(1). https://doi.org/10.1186/s12968-020-00619-x

MLA:

Wiesmüller, Marco, et al. "Cardiac T2 mapping: Robustness and homogeneity of standardized in-line analysis." Journal of Cardiovascular Magnetic Resonance 22.1 (2020).

BibTeX: Download