A COMPARISON BETWEEN 6-POINT DIXON MRI AND MR SPECTROSCOPY TO QUANTIFY MUSCLE FAT IN THE THIGH OF SUBJECTS WITH SARCOPENIA

Grimm A, Meyer H, Nickel MD, Nittka M, Raithel E, Chaudry O, Friedberger A, Uder M, Kemmler W, Engelke K, Quick HH (2019)


Publication Type: Journal article

Publication year: 2019

Journal

Book Volume: 8

Pages Range: 21-26

Journal Issue: 1

DOI: 10.14283/jfa.2018.16

Abstract

Background: Changes in muscle fat composition as for example observed in sarcopenia, affect physical performance and muscular function, like strength and power. Objectives: The purpose of this study was to compare 6-point Dixon magnetic resonance imaging and multi-echo magnetic resonance spectroscopy sequences to quantify muscle fat. Setting, participants and measurements: Two groups were recruited (G1: 23 healthy young men (28 +/- 4 years), G2: 56 men with sarcopenia (80 +/- 5 years)). Proton density fat fraction was measured with a 6-point product and a 6-point prototype Dixon sequence in the left thigh muscle and with a highspeed multi-echo T2*-corrected H1 magnetic resonance spectroscopy sequence within the semitendinosus muscle of the left thigh. To evaluate the comparability among the different methods, Bland-Altman and linear regression analyses of the proton density fat fraction results were performed. Results: Mean differences +/- 1.96 * standard deviation between spectroscopy and 6pt Dixon sequences were 1.9 +/- 3.3% and 1.5 +/- 3.6% for the product and prototype sequences, respectively. High correlations were measured between the proton density fat fraction results of the 6-point Dixon sequences and spectroscopy (R = 0.95 for the product sequence and R = 0.97 for the prototype sequence). Conclusions: Dixon imaging and spectroscopy sequences show comparable accuracy for fat measurements in the thigh. Spectroscopy is a local measurement, whereas Dixon sequences provide maps of the fat distribution. The high correlations of the 6-point Dixon sequences with spectroscopy support their clinical use. They provide higher spatial resolution than spectroscopy, but are not suitable for a more complicated spectral analysis to separate extra-and intramyocellular lipids.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Grimm, A., Meyer, H., Nickel, M.D., Nittka, M., Raithel, E., Chaudry, O.,... Quick, H.H. (2019). A COMPARISON BETWEEN 6-POINT DIXON MRI AND MR SPECTROSCOPY TO QUANTIFY MUSCLE FAT IN THE THIGH OF SUBJECTS WITH SARCOPENIA. Journal of Frailty & Aging, 8(1), 21-26. https://dx.doi.org/10.14283/jfa.2018.16

MLA:

Grimm, Alexandra, et al. "A COMPARISON BETWEEN 6-POINT DIXON MRI AND MR SPECTROSCOPY TO QUANTIFY MUSCLE FAT IN THE THIGH OF SUBJECTS WITH SARCOPENIA." Journal of Frailty & Aging 8.1 (2019): 21-26.

BibTeX: Download