Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol

Albrecht MH, Nance JW, Schoepf UJ, Jacobs BE, Bayer RR, Litwin SE, Reynolds MA, Otani K, Mangold S, Varga-Szemes A, De Santis D, Eid M, Apfaltrer G, Tesche C, Göller M, Vogl TJ, De Cecco CN (2018)


Publication Type: Journal article

Publication year: 2018

Journal

Book Volume: 28

Pages Range: 2134-2142

Journal Issue: 5

DOI: 10.1007/s00330-017-5150-z

Abstract

OBJECTIVES: To compare the diagnostic accuracy between low-kilovolt peak (kVp) (≤ 100) and high-kVp (> 100) third-generation dual-source coronary CT angiography (CCTA) using a kVp-tailored contrast media injection protocol. METHODS: One hundred twenty patients (mean age = 62.6 years, BMI = 29.0 kg/m2) who underwent catheter angiography and CCTA with automated kVp selection were separated into two cohorts (each n = 60, mean kVp = 84 and 117). Contrast media dose was tailored to the kVp level: 70 = 40 ml, 80 = 50 ml, 90 = 60 ml, 100 = 70 ml, 110 = 80 ml, and 120 = 90 ml. Contrast-to-noise ratio (CNR) was measured. Two observers evaluated image quality and the presence of significant coronary stenosis (> 50% luminal narrowing). RESULTS: Diagnostic accuracy (sensitivity/specificity) with ≤ 100 vs. > 100 kVp CCTA was comparable: per patient = 93.9/92.6% vs. 90.9/92.6%, per vessel = 91.5/97.8% vs. 94.0/96.8%, and per segment = 90.0/96.7% vs. 90.7/95.2% (all P > 0.64). CNR was similar (P > 0.18) in the low-kVp vs. high-kVp group (12.0 vs. 11.1), as ws subjective image quality (P = 0.38). Contrast media requirements were reduced by 38.1% in the low- vs. high-kVp cohort (53.6 vs. 86.6 ml, P < 0.001) and radiation dose by 59.6% (4.3 vs. 10.6 mSv, P < 0.001). CONCLUSIONS: Automated tube voltage selection with a tailored contrast media injection protocol allows CCTA to be performed at ≤ 100 kVp with substantial dose reductions and equivalent diagnostic accuracy for coronary stenosis detection compared to acquisitions at > 100 kVp. KEY POINTS: • Low-kVp coronary CT angiography (CCTA) enables reduced contrast and radiation dose. • Diagnostic accuracy is comparable between ≤ 100 and > 100 kVp CCTA. • Image quality is similar for low- and high-kVp CCTA. • Low-kVp image acquisition is facilitated by automated tube voltage selection. • Tailoring contrast injection protocols to the automatically selected kVp-level is feasible.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Albrecht, M.H., Nance, J.W., Schoepf, U.J., Jacobs, B.E., Bayer, R.R., Litwin, S.E.,... De Cecco, C.N. (2018). Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol. European Radiology, 28(5), 2134-2142. https://dx.doi.org/10.1007/s00330-017-5150-z

MLA:

Albrecht, Moritz H., et al. "Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol." European Radiology 28.5 (2018): 2134-2142.

BibTeX: Download