Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study

Van Hamersvelt RW, Schilham AMR, Engelke K, Den Harder AM, De Keizer B, Verhaar HJ, Leiner T, De Jong PA, Willemink MJ (2017)


Publication Type: Journal article

Publication year: 2017

Journal

Book Volume: 27

Pages Range: 4351-4359

Journal Issue: 10

DOI: 10.1007/s00330-017-4801-4

Abstract

OBJECTIVES: To investigate the accuracy of bone mineral density (BMD) quantification using dual-layer spectral detector CT (SDCT) at various scan protocols. METHODS: Two validated anthropomorphic phantoms containing inserts of 50-200 mg/cm3calcium hydroxyapatite (HA) were scanned using a 64-slice SDCT scanner at various acquisition protocols (120 and 140 kVp, and 50, 100 and 200 mAs). Regions of interest (ROIs) were placed in each insert and mean attenuation profiles at monochromatic energy levels (90-200 keV) were constructed. These profiles were fitted to attenuation profiles of pure HA and water to calculate HA concentrations. For comparison, one phantom was scanned using dual energy X-ray absorptiometry (DXA). RESULTS: At both 120 and 140 kVp, excellent correlations (R = 0.97, P < 0.001) were found between true and measured HA concentrations. Mean error for all measurements at 120 kVp was -5.6 ± 5.7 mg/cm3(-3.6 ± 3.2%) and at 140 kVp -2.4 ± 3.7 mg/cm3(-0.8 ± 2.8%). Mean measurement errors were smaller than 6% for all acquisition protocols. Strong linear correlations (R2 ≥ 0.970, P < 0.001) with DXA were found. CONCLUSIONS: SDCT allows for accurate BMD quantification and potentially opens up the possibility for osteoporosis evaluation and opportunistic screening in patients undergoing SDCT for other clinical indications. However, patient studies are needed to extend and translate our findings. KEY POINTS: • Dual-layer spectral detector CT allows for accurate bone mineral density quantification. • BMD measurements on SDCT are strongly linearly correlated to DXA. • SDCT, acquired for several indications, may allow for evaluation of osteoporosis. • This potentially opens up the possibility for opportunistic osteoporosis screening.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Van Hamersvelt, R.W., Schilham, A.M.R., Engelke, K., Den Harder, A.M., De Keizer, B., Verhaar, H.J.,... Willemink, M.J. (2017). Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study. European Radiology, 27(10), 4351-4359. https://doi.org/10.1007/s00330-017-4801-4

MLA:

Van Hamersvelt, Robbert W., et al. "Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study." European Radiology 27.10 (2017): 4351-4359.

BibTeX: Download