Contactless Carotid Pulse Measurement Using Continuous Wave Radar

Beitrag bei einer Tagung
(Konferenzbeitrag)


Details zur Publikation

Autor(en): Shi K, Schellenberger S, Steigleder T, Michler F, Lurz F, Weigel R, Kölpin A
Jahr der Veröffentlichung: 2018
Sprache: Englisch


Abstract

Cardiovascular diseases are one of the major causes of death. Regular
checkups and preventive actions can drastically help reducing fatal
incidences. This can be achieved by monitoring the carotid artery or
rather the carotid pulse signal. Commonly, ultrasound devices are used
for that purpose. However, these devices are costly, mostly stationary
and their usage requires training and experience. This paper
investigates the possible usage of radar systems as a contactless and
low-cost alternative for carotid pulse measurements. Theoretical
investigations reveal a linear relationship between the measurands of
both devices and synchronous recordings from three test persons further
confirm the feasibility of using radar systems as a potential device for
monitoring cardiovascular diseases.


FAU-Autoren / FAU-Herausgeber

Lurz, Fabian
Lehrstuhl für Technische Elektronik
Michler, Fabian
Lehrstuhl für Technische Elektronik
Schellenberger, Sven
Lehrstuhl für Technische Elektronik
Shi, Kilin
Lehrstuhl für Technische Elektronik
Weigel, Robert Prof. Dr.-Ing.
Lehrstuhl für Technische Elektronik


Autor(en) der externen Einrichtung(en)
Brandenburgische Technische Universität Cottbus-Senftenberg


Zitierweisen

APA:
Shi, K., Schellenberger, S., Steigleder, T., Michler, F., Lurz, F., Weigel, R., & Kölpin, A. (2018). Contactless Carotid Pulse Measurement Using Continuous Wave Radar. Kyoto, JP.

MLA:
Shi, Kilin, et al. "Contactless Carotid Pulse Measurement Using Continuous Wave Radar." Proceedings of the 2018 Asia-Pacific Microwave Conference, Kyoto 2018.

BibTeX: 

Zuletzt aktualisiert 2019-05-02 um 09:38