Extended ellipse-line-ellipse trajectory for long-object cone-beam imaging with a mounted C-arm system

Yu Z, Lauritsch G, Dennerlein F, Mao Y, Hornegger J, Noo F (2016)


Publication Status: Published

Publication Type: Journal article

Publication year: 2016

Journal

Publisher: Institute of Physics: Hybrid Open Access

Book Volume: 61

Pages Range: 1829-1851

Journal Issue: 4

DOI: 10.1088/0031-9155/61/4/1829

Abstract

Recent reports show that three-dimensional cone-beam (CB) imaging with a floor-mounted (or ceiling-mounted) C-arm system has become a valuable tool in interventional radiology. Currently, a circular short scan is used for data acquisition, which inevitably yields CB artifacts and a short coverage in the direction of the patient table. To overcome these two limitations, a more sophisticated data acquisition geometry is needed. This geometry should be complete in terms of Tuy's condition and should allow continuous scanning, while being compatible with the mechanical constraints of mounted C-arm systems. Additionally, the geometry should allow accurate image reconstruction from truncated data. One way to ensure such a feature is to adopt a trajectory that provides full R-line coverage within the field-of-view (FOV). An R-line is any segment of line that connects two points on a source trajectory, and the R-line coverage is the set of points that belong to an R-line. In this work, we propose a novel geometry called the extended ellipse-line-ellipse (ELE) for long-object imaging with a mounted C-arm system. This trajectory is built from modules consisting of two elliptical arcs connected by a line. We demonstrate that the extended ELE can be configured in many ways so that full R-line coverage is guaranteed. Both tight and relaxed parametric settings are presented. All results are supported by extensive mathematical proofs provided in appendices. Our findings make the extended ELE trajectory attractive for axially-extended FOV imaging in interventional radiology.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Yu, Z., Lauritsch, G., Dennerlein, F., Mao, Y., Hornegger, J., & Noo, F. (2016). Extended ellipse-line-ellipse trajectory for long-object cone-beam imaging with a mounted C-arm system. Physics in Medicine and Biology, 61(4), 1829-1851. https://dx.doi.org/10.1088/0031-9155/61/4/1829

MLA:

Yu, Zhicong, et al. "Extended ellipse-line-ellipse trajectory for long-object cone-beam imaging with a mounted C-arm system." Physics in Medicine and Biology 61.4 (2016): 1829-1851.

BibTeX: Download