Isotropic reconstruction of SPECT data using OSEM3D: Correlation with CT

Römer W, Reichel N, Vija H, Nickel I, Hornegger J, Bautz W, Kuwert T (2006)


Publication Type: Journal article, Original article

Publication year: 2006

Journal

Original Authors: Römer W., Reichel N., Vija H., Nickel I., Hornegger J., Bautz W., Kuwert T.

Publisher: Elsevier

Book Volume: 13

Pages Range: 496-502

Journal Issue: 4

DOI: 10.1016/j.acra.2005.12.004

Abstract

Rationale and Objectives. A statistical based iterative single-photon emission-computed tomography (SPECT) reconstruction algorithm (OSEM) modeling the depth-dependent collimator response in three dimensions has recently been introduced (OSEM3D). The aim of this study was to evaluate the axial shape fidelity of OSEM3D in comparison to OSEM, not taking this variable into account (OSEM2D). Materials and Methods. SPECT and separate spiral CT were performed in a phantom containing spheres filled with In-111. In-111-pentetreotide-SPECT and separate spiral-CT imaging were also performed in 22 patients with neuroendocrine tumors. Using window settings adapting the transversal size of the SPECT hot spots to that on CT and the 50% isocontour as boundary, the three-dimensional extensions (dx, dy, dz) of the SPECT representation of the structures under study were measured. These variables were also determined for CT. Furthermore, an index of eccentricity was calculated by averaging the ratios between dz and dx and dz and dy (IE). For isotropically imaged spheres, IE is 1. Results. For OSEM2D, IE was significantly different from 1 in the phantom data (P < .05); this was not the case for OSEM3D and CT. This finding was accounted for by a significantly greater dz on the OSEM2D-SPECT images. In the patient data, dz was by approximately 15.5% greater for OSEM2D than for the other two modalities (P < .05). Conclusions. The use of OSEM3D avoids deformation of hot SPECT lesions in z-direction. This may be of particular importance in SPECT/CT hybrid imaging capitalizing on the exact match of both modalities. © AUR, 2006.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Römer, W., Reichel, N., Vija, H., Nickel, I., Hornegger, J., Bautz, W., & Kuwert, T. (2006). Isotropic reconstruction of SPECT data using OSEM3D: Correlation with CT. Academic Radiology, 13(4), 496-502. https://doi.org/10.1016/j.acra.2005.12.004

MLA:

Römer, Wolfgang, et al. "Isotropic reconstruction of SPECT data using OSEM3D: Correlation with CT." Academic Radiology 13.4 (2006): 496-502.

BibTeX: Download