Radiation dose reduction in parasinus CT by spectral shaping

May M, Brand M, Lell M, Sedlmair M, Allmendinger T, Uder M, Wüst W (2017)


Publication Type: Journal article

Publication year: 2017

Journal

Book Volume: 59

Pages Range: 169-176

Journal Issue: 2

DOI: 10.1007/s00234-016-1780-0

Abstract

Spectral shaping aims to narrow the X-ray spectrum of clinical CT. The aim of this study was to determine the image quality and the extent of radiation dose reduction that can be achieved by tin prefiltration for parasinus CT.All scans were performed with a third generation dual-source CT scanner. A study protocol was designed using 100 kV tube voltage with tin prefiltration (200 mAs) that provides image noise levels comparable to a low-dose reference protocol using 100 kV without spectral shaping (25 mAs). One hundred consecutive patients were prospectively enrolled and randomly assigned to the study or control group. All patients signed written informed consent. The study protocol was approved by the local Institutional Review Board and applies to the HIPAA. Subjective and objective image quality (attenuation values, image noise, and contrast-to-noise ratio (CNR)) were assessed. Radiation exposure was assessed as volumetric CT dose index, and effective dose was estimated. Mann-Whitney U test was performed for radiation exposure and for image noise comparison.All scans were of diagnostic image quality. Image noise in air, in the retrobulbar fat, and in the eye globe was comparable between both groups (all p > 0.05). CNReye globe/air did not differ significantly between both groups (p = 0.7). Radiation exposure (1.7 vs. 2.1 mGy, p < 0.01) and effective dose (0.055 vs. 0.066 mSv, p < 0.01) were significantly reduced in the study group.Radiation dose can be further reduced by 17% for low-dose parasinus CT by tin prefiltration maintaining diagnostic image quality.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

May, M., Brand, M., Lell, M., Sedlmair, M., Allmendinger, T., Uder, M., & Wüst, W. (2017). Radiation dose reduction in parasinus CT by spectral shaping. Neuroradiology, 59(2), 169-176. https://dx.doi.org/10.1007/s00234-016-1780-0

MLA:

May, Matthias, et al. "Radiation dose reduction in parasinus CT by spectral shaping." Neuroradiology 59.2 (2017): 169-176.

BibTeX: Download